PRODUCT


    Features and Applications

    NEG pump is a kind of chemisorption pump, which assembled after the NEG alloy heated by high sintering, It could eliminates large amounts of residual gases in a vacuum environment, mainly applied for UHV testing or Lab equipment. When it’s activated NEG pumps could operating without power, also free of Vibration and Nonmagnetic. The highlight of the NEG pumps it's extremely effective for Hydrogen and other active gases, and would never reduce under the UHV.

    Basic Characteristics and General Data

    Type

    Length(mm)

    Outside Diameter (mm)

    Gross Weight(g)

    Ordinary Temperature Resistance(Ω)

    NP-172-40-S/I/16

    133±20

    25±3

    140±15

    0.28±0.05

    NP-172-60-S/I/35

    133±20

    25/36±3

    170±20

    0.28±0.05

    NP-175-80-S/I/35

    133±20

    25/36±3

    170±20

    0.28±0.05

    NP-172-120-S/I/35

    133±20

    25/36±3

    170±20

    0.28±0.05

    NP-172-720-S/I/100

    145±20

    80/95±3

    980±50

    2.1±0.15

    NP-T-400-C/0/150

    58±5

    145±15

    480±30

    /

    Type

    Gettering Rate(H2)

    Total Sorption Capacity(H2)

    Gettering Rate (CO)

    Total Sorption Capacity (CO)

    L/S

    Torr·L

    L/S

    Torr·L

    NP-172-40-S/I/16

    100

    600

    50

    300

    NP-172-60-S/I/35

    200

    1000

    100

    500

    NP-175-80-S/I/35

    200

    1000

    100

    500

    NP-172-120-S/I/35

    400

    2000

    210

    1000

    NP-172-720-S/I/100

    2400

    12000

    1260

    6000

    NP-T-400-C/0/150

    /

    /

    /

    /


    Suggested Activation Conditions

    It is recommended that the user use a constant current power supply to energize and activate the NEG Pump. Recommended activation conditions: Energizing activation at 500°C for 45min, the vacuum degree of the system in the whole process of activation should be better than 0.01Pa. Proper extension of the time will facilitate full activation of the NEG Pump. If the standard activation temperature cannot be reached, the activation time must be extended to compensate. It is necessary to ensure the vacuum degree of the vacuum chamber of the activation process, If the vacuum is too low, there will be hot wire corrosion, heater sputtering, suction material pollution, abnormal activation temperature and other adverse conditions.
    The NEG Pump releases a certain amount of gas during activation, in order to ensure the vacuum degree during the activation of the NEG Pump. We recommend that the NEG Pump should be activated under dynamic vacuum, and the activation process should be slowly and gradually increased from 1.5A until the predetermined current value is reached, the rapid deflation and the change of electrical parameters caused by the rapid change of the temperature of the NEG Pump must be avoided.

    Type

    Recommended Activation Current (A)

    Recommended Activation Power (W)

    NP-172-40-S/I/16

    4

    40

    NP-172-60-S/I/35

    4

    35

    NP-175-80-S/I/35

    5.2

    60

    NP-172-120-S/I/35

    5.2

    60

    NP-172-720-S/I/100

    5

    300

    NP-T-400-C/0/150

    /

    /

    Caution

    ●When activated and working, the NEG Pump casing and flange have a high temperature, pay attention to prevent burns.
    ●When the NEG Pump is at high temperature, it must be in vacuum conditions to avoid failures due to contamination and consumption.
    ●When connecting the power supply, ensure that the connection between the power supply and the flange electrode is firm, and pay attention to the insulation with other parts.
    ●Before heating activation, pay attention to ensure that the system is under vacuum conditions that meet the requirements.
    ●Under special circumstances, in order to make the NEG Pump have a high pumping speed for C, N, O and other gases, the working temperature can be maintained in the range of 200 °C ~ 250 °C (energized 3A), at which time the ultimate vacuum degree that the NEG Pump can achieve is reduced.
Chinese にほんご

Share

QQ
TEL

86-13512546708